CAUTION
· Do not autoclave the handpiece with other instruments even when it is in a pouch. This is to prevent possible discolouration and damage to the
handpiece from chemical residue on other instruments.
· Keep the handpiece in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The air should be free from dust, salt
and sulphur.
· Immediately after use, the handpiece should be cleaned, lubricated and sterilized. If blood remains on the external or internal surfaces it can
become clotted and cause rust.
· Do not heat or cool the handpiece too quickly. Rapid change in temperature could cause damage to the handpiece.
· If the sterilizer chamber temperature may exceed 135˚C during the dry cycle then delete the dry cycle.
· Autoclave sterilization is recommended for the product. The validity of other sterilization methods is not confirmed.
· Do not touch the handpiece immediately after autoclaving as it will be very hot and must remain in a sterile condition.
NOTICE
NSK recommends Class B sterilizers as stated in EN13060.