Precaution for use
The product may be used only after the user completely understands the proper methods
1
of use.
The Fixture may be damaged if the size of the surgical tool is incorrectly selected. Ac-
2
cordingly, the size of the screw to be removed and the connection to the fixture, among
others, must be verified and the appropriate Guide, Reverse Drill, and Screw Remover
must be selected.
If the Perfect Guide does not connect easily, a Conical, Internal, or External Guide that is
3
appropriate for the Fixture connection must be used. If use of the Perfect Guide is forced,
the fixture may become damaged. (Perfect Guide recommended connection torque: 15
Ncm)
The Conical, Internal, or External Guide must be affixed to the SR Driver Holder tightly
4
before use.
After attaching a guide, it is possible that the Screw can be removed just by reverse drilling.
5
Accordingly, drilling must be conducted with care.
The Reverse Drill and the Screw Remover must use the appropriate rotation speed (rpm).
6
(Reverse Drill = 2,000 rpm, Screw Remover = 80 rpm or lower)
The Reverse Drill is a single-use item, it cannot be reused.
7
How to Sterilize
Because the product is a non-sterilized medical device, select either a pre-vacuum or a
1
gravity autoclave. (Plastic products must not be sterilized at or above 170°C (338°F)
Before sterilization, the inner wrapper must be removed from the tray. Assembled compo-
2
nents must be separated in order to improve the efficiency of sterilization.
Using surgical wrap, wrap the tray, seal with autoclave tape, and sterilize.
3
<Recommended Steam Sterilization Conditions >
KIT,
Pre-vacuum¹²
Instrument
KIT,
Instrument
In order to effectively carry out high-pressure steam sterilization, the use of biological indicators at a regular
interval must be considered. (Dry heat sterilization or chemical sterilization is not recommended.)
① Minimum time and temperature conditions for steam sterilization to reach the sterilization guarantee level
-6
of 10
② If regional or national sterilization requirements are stricter than the conditions provided above, they must
be followed.
If the above sterilization conditions are exceeded, it is possible that the plastic and components may be
damaged. The sterilization device must be adjusted to ensure that the recommended temperatures are
not exceeded.
Cycle Type
Temperature
132 ℃
270 °F
121 ℃
Gravity¹
250 °F
Exposure
Pressure
Time
2 bars
3 minutes
28.5 psi
1 bars
40 minutes
14.5 psi
Rev.6 12/2015
Dry Time
30 minutes
30 minutes
NB-SRK-IFU-001