Stryker T/PUMP Serie Manual Del Operador página 18

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Problem
Warning indicator on with
unit in Standby mode.
Flow indicator and Standby
indicator are on with
T/Pump not pumping.
Temperature Setpoint light
blinking.
1-16
Troubleshooting
Possible Cause
Unit shut down in an over temperature
condition.
Unit detected a Flow warning for more
than 5 minutes, thus goes to standby.
Unit is warming up to the selected
setpoint.
Unit is in Cooling mode, for longer than
40 minutes.
101515 REV C
Remedy
Empty the reservoir and refill with room
temperature water.
Make sure all clamps are open.
Press the On/Standby button.
Verify flow through the pad.
The Warning light will turn off within 5
minutes.
Reference "Flow indicator light is on"
above.
Correct the problem, and press the On/
Standby to put the unit back into Run
mode.
Indication only.
Follow the shutdown procedure. Drain the
water in reservoir to ice fill level, and refill
with ice. Follow the start-up procedure.
www.stryker.com

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

T/pump tp600T/pump tp700

Tabla de contenido