Problem
Both the Temperature and
Therapy Cycle Setpoint lights
are blinking (TP700 Only).
T/Pump will not heat.
T/Pump will not cool.
Temperature or Therapy Time
buttons do not work (TP700
Only).
Water leaks from hose
connectors.
www.stryker.com
Troubleshooting
Possible Cause
Unit is in "Off" Therapy cycle time.
Reservoir is empty.
Flow is blocked.
Reservoir is empty.
Flow is blocked.
Ice is depleted.
The buttons have been locked.
Damaged O-ring.
Locking ring on Clik–Tite
connector is
®
not snapped into place (See Figure 5 on
page
1-11)
Quick disconnect not seated properly.
101515 REV C
Remedy
Indication only.
Refill with room temperature water to
proper level.
Reference "Flow indicator light is on"
above.
Refill with room temperature water to
proper level.
Reference "Flow indicator light is on"
above.
Drain excess water to Cooling water
line and fill remainder of reservoir with
ice.
Press and hold the lock button for two
seconds.
Replace Clik–Tite
®
connector.
Snap Clik–Tite
connector shut.
®
Secure pad connection to pump (See
Colder on
page
1-12) Replace connectors or pad
if defective.
1-17