8. Troubleshooting
Alpha Active 3
The following table provides a troubleshooting guide for the
system
in the event of malfunction.
Indicator
Possible Cause
Remedy
LOW PRESSURE
The tube-set is not connected
Check the tube-set connector
properly.
and ensure it is securely fitted to
the pump.
CPR not fully closed.
Close CPR unit.
There is a leak in the system.
Call service engineer.
POWER FAILURE
Power has been removed from
Re-apply power or switch the
the pump.
pump off.
SERVICE
Pump has detected an internal
Switch the pump off and call
fault.
service engineer.
16