Recommended separation distances between portable
and mobile RF communications equipment and the pump
The pump is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the pump can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the pump as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2√P
0.12
0.38
1.2
2.8
12
21
m
80 MHz to 800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23