Error message
Effect
Temperature calibrati-
Temperature calibration value is
on error
out of range
Temperature
adjust-
Water in and water out sensor
ment error
temperature have big deviation
Inside
temperature
Analysis of internal PCB temp.
too high
Sensor.
Inside temperature reaches limit
value.
Flow rate out of range Water flow rate of cooling wa-
ter is out of range
Flow rate too low
Water flow rate of cooling wa-
ter is less than the minimum li-
mit value
Condenser overload
Cooling power limit value is ex-
ceeded
Distillation dry
There is no more solvent dis-
tilled
Lift error
Lift can not reach the end po-
sition
Internal communica-
Internal Communication Error
tion error
PC
communication
Error command received from
error
PC via RS 232 or USB interface
Heating bath commu-
Communication with heating
nication error
bath is not available
EEPROM error
Error while reading or writing
the EEPROM
Storage error
Read or write internal flash er-
ror
If the actions described fail to resolve the fault or another error message is displayed, then take one of the following steps:
- contact IKA
service department,
®
- send the device for repair, including a short description of the fault.
In accordance with IKA
®
warranty conditions, the warranty
period is 24 months. For claims under the warranty please contact
your local dealer. You may also send the machine directly to our
factory, enclosing the delivery invoice and giving reasons for the
claim. You will be liable for freight costs.
Cause
Wrong simulator resistance is
chosen
PCB is broken
The cooling water temperature
is changing
Temperature sensor is not cali-
brated
Temperatrue sensor is broken
Room temperature > 40°C
PCB is broken
Water flow rate is too high
Water flow sensor is damaged
Water flow rate is too low
Water flow sensor is damaged
Condenser limit value is too lo-
wer
Distilling is too fast
Distillation is completed
Lift is overloaded or lift is blo-
cked
Position sensor failure
Lift motor, cable or PCB is da-
maged
Internal communication error
LWS uses wrong device for RV
10 control
Wrong tools are used to com-
municate with RV 10 control
Unstable connection
Infrared interface is blocked
Heating bath is switched off or
detect error
Heating bath with infrared in-
terface is not present
PCB is damaged
EEPROM is defective
Read or write internal flash er-
ror
Warranty
The warranty does not cover worn out parts, nor does it apply
to faults resulting from improper use, insufficient care or
maintenance not carried out in accordance with the instructions
in this operating manual.
Note: For claims on glassware units, please contact your local
dealer and do not send the glassware back to our factory.
Correction
Check the simulator resistance
Contact service department
Check the chiller and wait until the chiller
reaches the set temperature
Make sure there is no distilling
Contact service department
Contact service department
Switch off device to cool down
Contact service department
Reduce the water flow rate
Contact service department
Increase the water flow rate
Contact service department
Increase the condenser overload limit
Increase the set vacuum
Stop the distilling
Check the lift
Contact service department
Contact service department
Switch device on/off, Contact service de-
partment
Check the LWS setting
Check the PC tools
Check the connection
Check and clean the infrared interface
Switch on heating bath
Check the heating bath
Contact service department
Contact service department
Contact service department