en - Correct application
Re-usable laboratory utensils can be reprocessed using this module
in a Miele laboratory washer-disinfector. Follow the instructions for
your washer-disinfector as well as the instructions of the laboratory
utensil manufacturer on how to reprocess their items by machine.
The A 303 injector module is set up for reprocessing pipettes.
The module can take up to 98 pipettes, maximum length 450 mm.
During reprocessing the body of volumetric pipettes must not touch
each other as this would reduce the capacity of the module.
Use the
Pipettes
programme.
To ensure adequate pressure is used, only one A 303 module may be
placed in an A 150 lower basket. The module must only be placed in
the right hand side. The second module connection can be used with
a different type of module.
When using this module do not use an upper basket.
Laboratory glassware and laboratory utensils for reprocessing are re‐
ferred to as the wash load if they are not more closely defined.

Queries and technical problems

If you have any queries or technical problems that you would like to
discuss please contact Miele. Contact details can be found at the
back of the Operating Instructions for your Washer-Disinfector or go
to www.miele-professional.com.