5.4.2 Platform not connected

Platform not connected
Please relieve the platform and then connect the cable
between the PC and the platform.
0.0 Nm
0.0 Nm
0.0 %
0.0 %
Platform not connected
Please relieve the platform and then connect the cable
between the PC and the platform.

5.4.3 Missing ground contact

Missing ground contact
A force sensor is not placed on the floor.
Please check the surface below the measurement plate.
0.0 Nm
0.0 Nm
0.0 %
0.0 %
Missing ground contact
A force sensor is not placed on the floor.
Please check the surface below the measurement plate.
84
Live view
Live view
OK
0.0 Nm
0.0 Nm
0.0 %
0.0 %
1
2
OK
Live view
Live view
OK
0.0 Nm
0.0 Nm
0.0 %
0.0 %
1
2
OK
Error message text:
: "Platform not connected" (force measurement plate)
: "Please remove any loads from the platform and
then connect the cable between the PC and the plat­
form."
: "OK"
Possible causes:
Magnetic USB plug has been disconnected
USB plug not connected to mini-computer
Mini-computer software has crashed
Defective cable
Defective force measurement plate
Electromagnetic interference
Measures:
► Carry out the appropriate measures depending on
the cause (for example connect USB plug; wait for
booting to finish; restart the mini-computer; have
defective components repaired by Ottobock Ser­
vice).
► In case of electromagnetic interference, disconnect
the USB plug connection for 15 seconds and then
connect it again. Restart the mini-computer if
necessary (disconnect the mains plug, wait 3
minutes and connect it again).
► Tap on "OK" x 1.
Error message text:
: "Missing ground contact"
: "A force sensor is not placed on the floor."
: "Please check the surface below the measurement
plate." – Apply the load to the force measurement plate
as evenly and consistently as possible. Avoid a point
load near the edge of the plate since this can also
cause this error message.
: "OK"
Possible causes:
Uneven floor
Defective sensor in the force measurement plate
Defective circuit board in the force measurement
plate
Measures:
► Carry out the appropriate measures depending on
the cause (for example level the force measurement
plate on an even floor; have defective components
repaired by Ottobock Service).
► 1x tap on "OK".