Setting up the Weighing Platform
Selecting the Site of Installation
The surface must be able to bear the weighing platform
➜
under maximum load at the points of support. At the
same time it should be so stable that no vibrations
arise during weighing. This is also to be observed
when installing the weighing platform in conveyor and
similar systems.
If possible, platform installation should be located in
➜
an area free from vibration.
Ambient Conditions
Observe the following ambient conditions:
➜
– No direct sunshine
– No permanent water on the load cell
(max. 120 min. per day)
– Absolute humidity shall stay below 30 g/m3
(environment)
– No excessive temperature fluctuations
– Temperature range –10 °C to +40 °C