4
TROUBLESHOOTING
Error
Problem
message
The device is not working
properly.
While the symbol
flashing, a new
measurement cannot be
started in the ear/temple
mode.
The device is displaying a
considerable change to the
room temperature.
Room temperature greater
than 40.0
Room temperature less than
o
10.0
C (50.0°F).
is
o
C (104.0°F).
Possible cause
Software error
The device is not
yet ready to take a
new measurement
The device was not
stored as instructed
- 43 -
Solution
Remove the battery, wait
one minute and then
replace the battery (sect.
3.1).
If the error message
appears again, contact
your retailer.
Wait until the symbol has
stopped flashing. When it
has stopped flashing,
then start the new
measurement.
Place the thermometer in
a room with a
temperature of between
10.0°C (50.0°F) and
40.0°C (104.0°F) for at
least 30 minutes.