Guidance and manufacturer's declaration – electromagnetic immunity
The THixyz series is intended for use in the electromagnetic environment specified below.
The customer or the user of the THixyz series should assure that it is used in such an
environment.
Immunity test IEC 60601
test level
3 Vrms
Conducted RF
150 kHz
IEC 61000-4-6
80 MHz
3 V/m
Radiated RF
80 MHz
IEC 61000-4-3
2,5 GHz
Compliance
level
Not
applicable
to
3 V/m
to
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the THixyz series, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1,2
d = 1,2
80 MHz to 800 MHz
d = 2,3
800 MHz to 2,5 GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, a
should be less than the compliance level in each
frequency range b.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
- 47 -