Calibration checks - Testing instructions
(for trained specialist staff only)
5. Testing for divergence of pressure
indication and density of pressure
circuit (setting time for cuff at least
120 seconds) can now be carried
out in the usual manner.
6. Switch off unit and reconnect the
cuff to the unit.
C) Safety seal
As a safeguard, the upper and
lower parts of the housing can be
joined
with
Alternatively, a hole drilled in the
bottom of the monitor for the con-
necting screw for the housing can
be sealed for safety.
a
safety
seal.
71