Descargar Imprimir esta página

NSK Varios 170 Manual De Operación página 9

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

9. Troubleshooting
When trouble is found, please check the followings prior to consulting your dealer.
Problem
Item to Check
The Tip does not
No / Poor
generate vibration, in
spite of depressing
vibration.
the Foot Control.
The Tip is bent
or broken.
The Tip is flying
away.
Noise from the
handpiece.
The handpiece
is overheating.
The water does
not reach to the
handpiece.
Remove the
handpiece from the
No / Weak water.
handpiece cord. Water
does not comes out or
weak from it.
No / Weak water form
the handpiece.
Water is leaking from
the joint between the
Water leakage.
handpiece and the
cord.
Tip oscillate, but
Handpiece LED turns
Handpiece
on and off.
LED does not
illuminate.
Tip oscillate, but
(Varios 170 LUX)
Handpiece LED does
not turn on.
* Repairs cannot be made by the customer.
The Tip is not tightened firmly.
Worn Tip.
Power has not been correctly
adjusted for the Tip.
Failure of vibrator in the handpiece. Contact dealer.*
Power has not been properly
adjusted for the Tip.
The Tip is not tightened firmly.
Power has not been properly
adjusted for the Tip.
The Tip is not tightened firmly.
Failure of vibration in the handpiece
or the Module.
Power has not been properly
adjusted for the Tip.
The Tip is not tightened firmly.
Failure of vibration in the handpiece
or the Module.
The water adjustment dial is
closed.
The water filter is clogged.
There may be exogenous material
in the hole.
O-Ring at the handpiece cord is
worn or damaged.
The handpiece is not connected into
the Handpiece Cord correctly.
Disconnection in the Handpiece
Cord, or failure in Module.
Cause
Tighten the Tip until the Tip Wrench clicks.
Replace the Tip.
Adjust the power on the Power Guide or
Tip case label. Do not exceed.
Adjust the power level the Power Guide or
Tip case label. Do not exceed.
Tighten the Tip until the Tip Wrench clicks.
Adjust the power level on the Power Guide
or Tip case label. Do not exceed.
Tighten the Tip until the Tip Wrench clicks.
Contact dealer.*
Adjust the power level on the Power Guide
or Tip case label. Do not exceed.
Tighten the Tip until the Tip Wrench clicks.
Contact dealer.*
Check the water circuitry and supply to the
Unit Chair.
Turn the water adjustment dial and set it to
an appropriate water colume.
Contact dealer.*
Air it with the appliances such as
syringes from the rear of the handpiece.
If it is still unsolved, contact dealer.
Replace with new O-Ring (Refer to 6 6-2
Changing O-Ring).
Firmly insert the handpiece into the
Handpiece Cord inmost.
Contact dealer.*
8
Solution

Publicidad

loading