Rs 232 Interface - Heidolph PD 5001 Manual De Instrucciones

Accionamiento de la bomba
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

IEI
TTL-level 0 V = CW rotation, TTL-level 5 V = CCW rotation.
-
A 5 volts TTL signal turns item ON, next signal to follow turns item OFF. This function is
-
available all the time (in internal mode either).
For triggering sense of rotation and ON / OFF function, a 5 volts pin is available in the
-
interface connector, which allows potential-free switching.
The analog interface is recognized at a level of 0.2 volts or a 4 m Amps current; This mode is
-
confirmed by "EA" in the display. As long as the pump drive is controlled via the analog
interface, Start / Stop-key will be the only active key on the item.
Wiring of the interface connector (37):
-
fig. 49
View on terminal end
Use IP 67 plug connectors only (Binder 6-pin 99-5121-15-06), Heidolph P/N 14-010-006-81(,
-
not to void splash water protection.
Installation of plug connectors is limited to skilled personnel.
-
9.4.2. RS 232 interface (fig. 50)
The RS 232 interface allows pump control from a PC via parallel interface.
-
The RS 232 interface connector is installed in the rear panel of the pump drive. Wiring:
-
Use IP 67 plug connectors only (Binder 8-pin 99-5171-15-08), Heidolph P/N 14-010-006-82,
-
not to void splash water protection.
Installation of plug connectors is limited to skilled personnel.
-
Interface parameters: 9600 Baud, no parity, 8 bit, 1 stop bit.
As long as the RS 232 interface is active, all of the pump's control elements and displays will
-
remain active.
PIN 1 = +5 volts
PIN 2 = start/stop
PIN 3 = 0-10 V
PIN 4 = sense of rotation
PIN 5 = 0 V (GND)
PIN 6 = 4-20 mA
PIN 1 = PIN 1
PIN 2 = PIN 6
PIN 3 = PIN 2
PIN 4 = PIN 7
PIN 5 = PIN 3
PIN 6 = PIN 8
PIN 7 = PIN 4
PIN 8 = PIN 5
View on terminal (soldered) end)
9-pin.SUB-D
66
fig. 50

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pd 5006Pd 5101Pd 5106Pd 5201Pd 5206

Tabla de contenido