An SD memory card socket is mounted to the PCB to support the optional SD
memory and bus connectors are included for the option boards.
Scale Bases
The IND246 terminal supports analog scale bases and provides either 10 volts
(AC version) or 5 volts (battery version) of excitation to drive analog load cells. Up
to four (battery version) or eight (AC version) 350Ω load cells can be powered by
the terminal.
A six wire load cell connection is provided with sense lines to help maintain
accuracy as the load cell cable resistance changes with temperature variations.
Options
The following options are available for the IND246:
COM2 Serial Port
•
COM2 and DIO (relay output)
•
USB Port
•
USB and DIO (relay output)
•
Ethernet Port
•
COM2 Serial Port
This optional port provides RS-232 and RS-485 communication at rates from 300
to 115.2k baud. The port is bidirectional and can be configured for various
functions such as demand output, continuous output, extended continuous output,
SICS host communications or ASCII command input (C, T, P, Z).
One RS-232/485 serial COM port
One RS-232/485 serial COM port
Internal, discrete I/O (2 inputs and 4 outputs)
─
Inputs are optically isolated solid state and switch selectable as either
active or passive
─
Output relays provide one normally open contact per relay
One USB 2.0 compliant port, hardware bridge. Acts as virtual (UCP) COM port
One USB 2.0 compliant port, hardware bridge. Acts as virtual (UCP) COM port
Internal, discrete I/O (2 inputs and 4 outputs)
─
Inputs are optically isolated solid state and switch selectable as either
active or passive
─
Output relays provide one normally open contact per relay
One 10/100 Ethernet port with automatic link polarity detection and correction.
Supports TCP/IP socket connection. Does not support FTP
IND246 Terminal Installation Manual