Descargar Imprimir esta página

Error Messages/Troubleshooting - Beurer medical BC 19 Manual De Instrucciones

Ocultar thumbs Ver también para BC 19:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 37
• In order to retrieve measurement results again, press
the memory key M. Initially, the average value of the
last three blood pressure readings is displayed (Dis-
play ). Press the memory key M again to display the
latest measurements with date, time and graphic WHO
classification and to hear the values announced. You
can adjust the volume during the announcement by
pressing the volume key + and -.
• You can clear the memory by pressing and holding +
and - simultaneously for 5 seconds after pressing the
memory key M and selecting the user.
• If you wish to change the user memory, see "Select
memory".

7. Error messages/troubleshooting

In case of faults, the
_ message appears in the display.
Error messages may appear if:
• systolic or diastolic pressure could not be measured
(
or
appears on the display)
• systolic or diastolic pressure was outside the measure-
ment range ( or Lo appears on the display)
• the cuff is fastened too tightly or loosely ( Er3 or Er4
appears on the display)
• the pump pressure is higher than 300 mmHg ( Er5
appears on the display)
• pumping up takes longer than 160 seconds ( Er6
appears on the display)
• there is a system or device error ( ErA ,
Er8 appears on the display)
• the batteries are almost empty
In the above cases, you must repeat the measurement.
Make sure that the cuff tube is properly inserted and that
you do not move or talk.
Re-insert the batteries if necessary, or else replace
them.
Technical alarm – description
Should the recorded blood pressure (systolic or diasto-
lic) lie outside the limits specified in the section "Techni-
cal specifications", the technical alarm will appear on the
display indicating either " " or " Lo ". In such cases, you
should seek medical assistance and check the accuracy
of your procedure.
The limit values for the technical alarm are factory set
and cannot be adjusted or deactivated. These alarm limit
values are accorded second priority under the standard
IEC 60601-1-8.
The technical alarm is a non-locking alarm and must not
be reset. The signal shown on the display will disappear
automatically after about 8 seconds.
22
, Er7 or
.

Publicidad

loading