NE-C801S-KDE(V)_main.book Page 18 Friday, December 14, 2018 10:32 AM
4.Cleaning and disinfecting
Parts
Disinfecting
Materials
method
Alcohol
Disinfecting
ethanol
Sodium
hypochlorite
Milton*
(0.1%, 15min.)
Quaternary
ammonium
A
Osvan*
(0.1%, 10min.)
Chlorhexidine
Hibitane*
(0.5%, 30min.)
Amphoteric
Surfactant
Tego 51*
(0.2%, 15min.)
B
Boiling
* an example of commercially available disinfectant. (The concentration and residence time
specified in the table are under the conditions where the service life of the parts is tested with
each disinfectant used as described in its instruction manual. Please note that the testing was
not performed for the purpose of ensuring the effectiveness of the disinfectants. There is no
intention of suggesting that such disinfectants be used. Conditions of use and ingredients of
disinfectant vary for manufacturers.
Please read the instruction manuals carefully before use and apply disinfectant to each part in
an appropriate way. Please notice that service life of the parts may be shorter depending on
conditions, environments and frequency of use.)
** Remove the rubber band from mask before disinfecting.
***The band is not made with natural rubber latex.
A. Use a commercially available disinfectant. Follow the instructions provided by the
disinfectant manufacturer.
Note: Never clean with benzene, thinner or a flammable chemical.
B. Parts may be boiled between 15 to 20 minutes.
After boiling, carefully remove the parts, shake off excess water
and allow to air dry in a clean environment.
18
o: applicable X: not applicable
o
o
o
o
**
o
o
o
o
**
o
o
o
o
**
o
o
o
o
**
o
o
o
o
**
X
o
o
o
X
X
X
X
X
o
X
X
o
X
X
o
X
X
o
X
X
X