English
Max./Min. Alert flickering
No matter the alarm is on or off (the
and yellow backlight flickers twice every second continuously while the pulse
rate exceeds the Max./Min. threshold.
Probe failure alarm
If the probe is not properly attached to the device or if the probe is not Solaris
(and/or Nellcor) compatible, the
second with alarm sounded continuously.
Battery
Low battery
When the battery is low,
appears on the screen. Please replace the batteries as
soon as possible.
When the battery is critically low, the
1Hz frequency with alarm sounded for 1 minute before the device is off
automatically.
When
or
is on the screen, the alarm is off and the
backlight flickers once every second for 1 minutes before the device is off
automatically.
5.3 Setup
Preview
Press the
button to enter the setup mode.
The Max. value of SpO
(with a default value of 100), the Min. value of SpO
2
a default value of 86), the Max. value of pulse rate (with a default value of 140),
the Min. value of pulse rate (with a default value of 45) and memory interval setup
(default value of 5 minutes) appear in that order at an interval of 2.5 seconds. The
device will then exit the setup mode 2.5 seconds later. The buzzer icon (
and date appears on the screen continuously.
Press the
button to exit the "setup" mode.
Max. SpO
setup
2
Press the
button for 2 seconds to enter the setup mode.
To setup of Max. SpO
, the
icon of SpO
2
the Max. SpO
appears. If no adjustment is needed, press
2
value of SpO
or wait 10 seconds to exit.
2
or
is on the screen), the
icon and yellow backlight flicker once every
icon and yellow backlight flickers at
and yellow
(with
2
or
flickers once every second and with
2
to setup the Min.
8
or
)