Use in Hazardous Areas

Standard versions of the IND131 and IND331 terminals are not certified for use within hazardous
(explosive) areas. Specially marked terminals (models IND131xx and IND331xx) are FM and
KEMA certified for use in hazardous areas classified as Division 2 or Zone 2/22.
1.5.

Inspection and Contents Checklist

Verify the contents and inspect the package immediately upon delivery. If the shipping container is
damaged, check for internal damage and file a freight claim with the carrier if necessary. If the
container is not damaged, remove the terminal from its protective package, noting how it was
packed, and inspect each component for damage.
If shipping the terminal is required, it is best to use the original shipping container. The terminal
must be packed correctly to ensure its safe transportation.
The package should include:
1-6
METTLER TOLEDO IND131/IND131x/IND331/IND331xx Installation Manual
Table 1-3: IND131, IND331 Power Consumption (DC Source)
Input Voltage
18VDC (min)
24 VDC
36 VDC (max)
Values shown are with internal COM2/DIO option and Analog Output option installed and
load cell input loaded with 4 x 350Ω load cells.
DO NOT USE THE STANDARD IND131 OR IND331 TERMINALS IN AREAS CLASSIFIED AS
HAZARDOUS BECAUSE OF COMBUSTIBLE OR EXPLOSIVE ATMOSPHERES. SPECIAL MODELS
(IND131xx AND IND331xx) ARE AVAILABLE FOR THESE APPLICATIONS. CONTACT AN
AUTHORIZED METTLER TOLEDO REPRESENTATIVE FOR INFORMATION ABOUT HAZARDOUS
AREA APPLICATIONS.
DIV 2 AND ZONE 2/22 INSTALLATION
IF YOU WISH TO INSTALL THE IND131xx OR IND331xx TERMINAL IN A DIVISION 2 OR ZONE
2/22 AREA, REFER TO THE DIVISION 2 AND ZONE 2/22 INSTALLATION INSTRUCTIONS
INCLUDED ON THE RESOURCE CD PROVIDED WITH THE TERMINAL. FAILURE TO COMPLY
WITH THE INSTRUCTIONS PROVIDED THERE COULD RESULT IN BODILY HARM AND/OR
PROPERTY DAMAGE.
Note: When an IND131xx or IND331xx terminal is installed into an area classified as Division 2 or zone
2/22, special AC wiring requirements must be met. Refer to document 64068795 Division 2 and Zone
2/22 Installation Guide.
IND131
IND331
I (mA)
P (W)
I (mA)
158
2.84
170
120
2.88
130
84
3.02
90
WARNING
P (W)
3.06
3.12
3.24
64067480
|
11
|
05/2018