To reduce the risk of injury from possible end user exposure associated
with a) biological contamination and b) degradation products released
from the purifi cation media; and
To reduce the risk of product loss associated with a) chemical release from
purifi cation media, and b) nullifi cation of Supplier's USP<88> Class VI
compliance of fi lter capsule components:
• Only expose the capsule to 1 sterilization or 1 sanitization cycle prior to use.
• Do not autoclave and base sanitize the same capsule prior to use.
• During pre-use autoclave sterilization do not exceed a temperature
of 121°C or a duration longer than 30 minutes for numbers BV0.3R
(EMP101AEX020R) and BV1R (EMP201AEX020R)
• Do not use with or expose this product to process liquids until capsule
temperature goes <40°C (104°F).
• Only use NaOH as sanitizing chemical and do not exceed 1M NaOH
concentration or 1-hour duration for sanitization.
• Use only 70% ethyl and/or isopropyl alcohol to wipe the exterior of the
capsules.
• Do not use with or expose this product to pressurized steam or sanitizing
chemicals other than NaOH.
WARNING
- 4 -