Chapter 4 Product Main Characteristics - Gima 33232 Manual De Uso

Ocultar thumbs Ver también para 33232:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

• If the defibrillation time is over 5 seconds, the one-off chest electrode should be
used to prevent the metal electrode from burning patients' skin. It is better not to
use this device with other electric stimulating devices simultaneously. If it is nec-
essary, there must be professional technician guiding on the scene.
• In the process of device operation, doctor should be in the operating place all the
time and observe the patient carefully. Cut off the power or take off the electrode
when necessary to ensure patient's security.
• After operation, turn all of the function state back to the original state before turn-
ing off. Take off the electrode gently without forcible pulling.
Clean device and all accessories for the next use.
• With over loading, amplifier saturation or lead off, device can prompt working fail-
ure.
• The device and all accessories should be performed maintenance and overhaul
termly (the period should be no longer than half a year).
• Electrocardiograph is a kind of measuring device, should be send to the related
department to test termly following local regulation. The test cycle should not be
more than 1 year.
Chapter 4
PRODUCT MAIN CHARACTERISTICS
4.1 High resolution thermal-array output system (8 dots/mm), needs no adjustment.
Frequency Response is up to 150Hz.
4.2 Record clear and exact three channels of ECG waveforms and remark real-time
and continuously. The remark includes: lead sign, sensitivity, paper speed, filter
state, etc.
4.3 In auto mode, one button operation to complete record, which will improve work
efficiency.
4.4 Key-press and touch-screen control, more convenient for operation.
TFT screen shows the working status, more clear for observation.
4.5 Safety Class: Class I, Type CF applied part with defibrillation protection.
4.6 The power supply includes both AC and DC. This device includes built-in lithium
polymer rechargeable battery.
If use power adapter for charging, about 4 hours is enough.
4.7 This instrument can record 450 pieces of ECG waveform and print for 4 hours
continually under the best DC state.
4.8 This device can print or store different time length of case to convenient for doc-
tor's case review and statistics.
4.9 You can select print mode, print length, gain, paper speed and report print selection.
4.10 You can select whether to set patient information and to save case when printing.
4.11 You can replay the stored case and patient information anytime.
4.12 Exterior SD card, convenient for use. With SD card, the device can store 100 case
most.
4.13 With USB2.0 port, working mode can be selected freely between store and syn-
chronization.
4.14 In the process of use, there are clear prompts (such as paper-lack, SD card oper-
ating error, etc.) for misoperation, which is convenient for users' operation.
ENGLISH
46

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Ecg90a

Tabla de contenido