J
Right
2 Product description
The product (3R15, 3R49) has the following key features:
•
Monocentric prosthetic knee joint
•
Product components for stance phase stability:
-
Adjustable, load-dependent braking mechanism to secure the
extension position
•
Product components for swing phase control:
-
Extensor (spring force adjustable)
•
Material: Stainless steel (3R15), Titanium (3R49)
3 Intended use
3.1 Indications for use
The product is intended exclusively for lower limb exoprosthetic fit
tings.
3.2 Area of application
CAUTION
Excessive strain on the product
Fall due to breakage of load-bearing components
► Only use the product according to its allowable field of applica
tion.
Allowable field of application (3R15)
Recommended mobility grade: 1 + 2
Allowable body weight: ≤ 100 kg
MG 1: indoor walker
The patient has the ability or the potential to use the prosthesis for
the purpose of moving or proceeding slowly on level floors. The
amount of time and the distance he/she can walk are seriously lim
ited due to his/her condition.
Wrong
Everyday prosthesis
MG 2: Restricted outdoor walker
The patient has the ability or the potential to walk slowly with the
prosthesis and to negotiate low environmental obstacles like curbs,
single steps or uneven surfaces. The amount of time and the dis
tance he/she can walk are seriously limited due to his/her condition.
Unallowable amputation types
Hip disarticulation, Hemipelvectomy
3.3 Combination possibilities
CAUTION
Failure to observe the manufacturer requirements regarding
combination possibilities
Injuries, malfunctions or product damage due to unallowable com
bination of prosthesis components
► Based on the instructions for use of all prosthetic components
used, verify that they may be combined with each other and are
approved for the patient's field of application.
► Contact the manufacturer with questions.
INFORMATION
In a prosthesis, all prosthetic components have to meet the patient's
requirements regarding the amputation level, body weight, activity
level, environmental conditions and field of application.
Unallowable combinations
Prosthetic hip joints
3.4 Environmental conditions
CAUTION
Use under unallowable environmental conditions
Fall due to damaged product
All reference numbers