Wash the components of the disassembled nebulizer,
the mouthpiece, the masks and nosepiece by using tap
water; dip in boiling water for 5 minutes.
Reassemble the nebulizer components and connect it
to the air-outlet (Pic. G), switch the device on and let it
work for 10-15 minutes.
!
Do not boil nor autoclave the air tube.
2. Disinfection:
Cold disinfecting solutions can be used, instead of boiling water,
according to manufacturer's instructions.
REPLACEMENT OF THE NEBULIZER
The nebulizer must be replaced after a long period of inacti-
vity, in case it shows deformations or breakings, or when the
nebulizer nozzle is obstructed by dry medicine, dust, etc. boso
recommends to replace the nebulizer after a period between
6 months and 1 year depending on the usage.
Use original nebulizers only.
REPLACEMENT OF THE AIR FILTER
Under normal usage conditions the air filter must
be replaced after approx. 500 operating hours or 12
months at the latest.
boso recommends to periodically check the air filter
(10 – 12 treatments) and, if the filter shows a grey or
brown colour or is wet, replace it. Extract the filter (Pic.
A-5) and replace it with a new one.