Virtual SVV™ Instructions for Use - EN
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Virtual SVV™ is intended for use in the electromagnetic environment specified below. The customer or the
user of Virtual SVV™ should assure that it is used in such an environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration — electromagnetic immunity
Virtual SVV™ is intended for use in the electromagnetic environment specified below. The customer or the
user of Virtual SVV™ should assure that it is used in such an environment,
Immunity test
Conducted RF
IEC / EN 61000-4-6
Radiated RF
IEC 60601 Test
level
+6 kV contact
+8 kV air
+2 kV for power supply
lines
+1 kV for input/output
lines
+1 kV differential mode
+2 kV common mode
< 5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
3 A/m
IEC / EN 60601
test level
3 Vrms
150kHz to 80 MHz
3 V/m
Compliance
+6 kV contact
+8 kV air
Not applicable
+1 kV for
input/output lines
Not applicable
Not applicable
3 A/m
Compliance level
3 Vrms
3 V/m
Page 43
Electromagnetic
Environment-Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be greater than
30%.
Mains power quality should be that
of a typical commercial or
residential environment.
Mains power quality should be that
of a typical commercial or
residential environment.
Mains power quality should be that
of a typical commercial or
residential environment. If the user
of Virtual SVV™ requires
continued operation during power
mains interruptions, it is
recommended that Virtual SVV™
be powered from an
uninterruptable power supply or its
battery.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or residential
environment.
Electromagnetic environment –
guidance
Portable and mobile RF
communications equipment should
be used no closer to any parts of
Virtual SVV™, including cables,
than the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation distance
=
d
2 , 1
P
=
d
2 , 1
P
80 MHz to 800 MHz