Analog Output

The Model 7500FO provides analog output signals for SpO
and event markers. Each output level conforms to the specifications
shown below:
Output
SpO
Output Analog Range
2
Pulse Rate Output Analog Range
Event Marker
Analog Output Load Current
%SpO
Analog Output Accuracy
2
Pulse Rate Analog Output Accuracy ±5%
Analog Output Calibration
Analog calibration signals that allow external device calibration are
provided after initial power up, and continue until the Model 7500FO
begins tracking SpO
ends when the system begins tracking signals. The calibration signal
sequence is as follows:
26
0 - 1.0 VDC (representing 0-100%)
1.27 VDC (out of track)
0 - 1.0 VDC (representing 0-300 BPM)
1.27 VDC (out of track)
0 VDC or 1.0 VDC nominal (representing
an event).
Event marker high for SpO
alarm limit.
2 mA maximum
±2%
and pulse rate readings. The calibration routine
2
Time Interval
30 seconds
30 seconds
1 second
1 second
1 second
1 second
1 second
1 second
1 second
1 second
1 second
1 second
1 second
Repeat
Memory and Data Output Features
Specification
Analog Signal
1.0 VDC
0.0 VDC
0.1 VDC
0.2 VDC
0.3 VDC
0.4 VDC
0.5 VDC
0.6 VDC
0.7 VDC
0.8 VDC
0.9 VDC
1.0 VDC
1.27 VDC
, pulse rate,
2
less than low