Steinberg SBS-KW-1000 Serie Manual De Usuario página 21

Ocultar thumbs Ver también para SBS-KW-1000 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 24
3) CALIBRATION
To meet metrology government's requirement in some of European countries, the
scale is protected against a user calibration by a calibration plug. In that case,
the user must take off the front panel and short-cut the calibration plug P2. (see
pic. above).
EN
If the calibration plug P2 is not short-cut, the error message
displayed, indicating there is no plug on the required position.
After entering the Calibration settings, the plug P2 should not be shorted anymo-
re, otherwise, the scale will auto power-off.
This does not apply to crane scale SBS-KW-1000/10B.
Entering Calibration Settings
 Press
twice to enter the Password mode.
The scale displays
. Press
or
, and
, to move the digital. Input the Calibration password: P8416.
Press
to confirm the input and to enter Calibration.
The scale displays
Press
to enter the System Unit settings.
System Unit
The Scale displays the System Unit, e.g.
is kg. The System Unit can be set to: kg and lb.
 Press
or
, and
 Press
to exit and save. Press
 Press
to enter the Calibration Gravity settings.
or
, and
to modify the digital. Press
, indicating that the system unit
, to change the System Unit.
or
to exit without saving.
37
CALIBRATION GRAVITY
The scale displays the Calibration Gravity, e.g.
Gravity g = 9.794. The Calibration gravity range can be set from 9,700 to 9,899.
 Press
to move the digital. Input the Calibration Gravity value.
 Press
 Press
MAX. CAPACITY
The scale displays the Max. Cap., e.g.
-Do NOT attempt to set Max. Cap. greater than scale's actual ca-
pacity. Overloading causes severe harm to scale, and is very dangerous.
 Press
 Press
 Press
ZERO DETECTION
The scale displays
Keep the scale without load and press
The message
will be
matically enter the Load1 Detection.
In most cases, a one load calibration is enough. Calibrating scale with
more than one weight is usually required only when the scale's linearity
performance is not desired.
LOAD 1 DETECTION
The scale displays
Load a standard weight, and press
The scale displays "00000", waiting for user input.
Press
and then
The weight must exceed 0, otherwise
The weight must be lighter than 100% F.S., otherwise
If the weight is equal to 100% F.S., the scale automatically saves and exits
Calibration.
It is recommended to use a weight that is equal to the scale's Max. Cap. to
calibrate the scale.
Rev. II.2017
Rev. II.2017
or
, and
to modify the digital. Press
to exit and save. Press
or
to enter the Max. Cap. settings.
,indicating that the Max. Cap. is 3t.
or
, and
, to change the Max. Cap.
to exit and save. Press
or
to enter the Zero Detection settings.
.
to start the weight detection.
displays. After the detection is finished, the scale auto-
.
to go to weight input interface.
or
, and then
to modify the digital. Press
, to move the digital. Input the Load1 weight.
displays.
38
, indicating the Calibration
or
, and
to exit without saving.
to exit without saving.
EN
or
,
displays.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Sbs-kw-1000bSbs-kw-1000rSbs-kw-1000ySbs-kw-1000/10b

Tabla de contenido