Troubleshooting

Trouble
Key switch does nothing
No correspondence between
displayed energy and effects
on skin
Simmer failure

Error Messages

Error Message
Low coolant flow
Over temperature
Shutter failure
No fiber attached
Remote open
Pedal not connected
No energy
Cook Medical © 2012
Possible Cause
• Power cord not connected
• Main line breaker off
• Emergency stop button
pressed
• Damaged fiber
• Damage blast shield
• Flash lamp not functional
Possible Cause
• Not enough water
• Room temperature too high
• Airflow is obstructed
• Cooling system component
failure
• Component failure
• Incorrect installation of
the fiber
• Wrong fiber installed
• The remote interlock is not
connected or is opened
• The remote interlock is
damaged
• The foot switch is not
connected
• The foot switch is damaged
• Optical component damage
Return to Table of Contents
Maintenance
Solution
• Connect power cord
• Check MAINS breaker
• Release the emergency stop
button
• Change fiber
• Change blast shield
• Call for service
Solution
• Fill water reservoir
• Lower the room temperature
• Ensure nothing is obstructing
the fan
• Call for service
• Call for service
• Check the proper connection
of the fiber
• Install suitable H-30 fiber
• Check the remote interlock
connection
• Call for service
• Connect the foot switch
properly
• Call for service
• Call for service
Chapter 6
1-35