When the Request is "E", the Answer will contain the
complete set of the following parameters:
Bytes
0000=STOP
0001=WAITING INTERLOCK
LSB X
0010= STARTING
X
0011 = NORMAL OPERATION
X
0100= HIGH LOAD
X
0101= ""
1
0110 =FAILURE
0111=APPROACHING LOW SPEED
X
...
X
...
X
...
MSB X
...
2÷5
Cycle Time
Pump Life
6÷9
10÷11
Pump Temperature
12
Power*
13
Voltage*
Frequency
14÷17
18÷19
Cycle #
20
R1 Status
21
R2 Status
22
CRC
*The values for power and voltage are given as numbers,
scaled from 0 to 255, where 0 corresponds to 0 V and 255
to the full scale voltage (100 V) or power (500 W).
When the Request is "G", the Answer will contain a string
of 12 characters with the following parameters:
Bytes
1
Speed adjust value (in KRPM)
2-3
Pump cycles number (integer coded in 2 bytes)
4
Speed threshold
5-8
Run up time in seconds
(long coded in 4 bytes)
9
Deat time
(0 = NO 1 = YES)
10
Soft Start mode
(0 = NO 1=YES)
11
Water cooling
(0 = NO 1=YES)
12
CRC
Meaning
""
Meaning
When the configuration parameters have to be changed,
send a Request string with "H" characters of the following
type:
"H" + .... DATA .... + CRC
The CRC is calculated on the string consisting of
H + DATA.
The following parameters can be changed:
Bytes
1
Speed adjust value (in KRPM)
2
Speed threshold
3-6
Run up time in seconds
(long coded in 4 bytes)
7
Deat time
(0 = NO 1 = YES)
8
Soft Start mode
(0 = NO I=YES)
9
Water cooling
(0 = NO I=YES)
10
CRC
When the Request is "I" (Status readings), the Answer
will contain a string of 2 characters with the following
parameters:
Bytes
0000 = STOP
0001 = WAITING INTERLOCK
LSB X
0010 = STARTING
X
0011 = NORMAL OPERATION
X
0100 = HIGH LOAD
0101 = " "
X
0110 = FAILURE
0111 = APPROACHING LOW SPEED
1
X
R2 status (0 = OFF 1 = ON)
X
R1 status (0 = OFF 1 = ON)
X
...
MSB X
...
2
CRC
When the Request is "J" (Numerical readings), the
Answer will contain a string of 5 characters with the
following parameters:
Bytes
1
Power (0-255 scaled)
2
Voltage (0-255 scaled)
3
Rotational speed KRPM
4
Pump temperature °C
(0-70 temperature reading, 255 = fail)
5
CRC
58
TECHNICAL INFORMATION
Meaning
Meaning
" "