Gima 32926 Manual Del Usuario página 43

Ocultar thumbs Ver también para 32926:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

43
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be
used in such environments:
IMMUNITY TEST
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies..
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
IEC 60601
COMPLIANCE
TEST LEvEL
LEvEL
3 Vrms
Not Applicable
150 kHz
to 80 MHz
3 V/m
3 V/m
80 MHz
to 2.5 GHz
ENGLISH
ELECTROMAGNETIC
ENvIRONMENT – GUIDANCE
Portable and mobile RF communications
equipment should be used no closer to
any part of the device, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2 80 MHz to 800 MHz
d = 2.3 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity
of equipment marked with the following
symbol:
a
b

Publicidad

Tabla de contenido
loading

Tabla de contenido