21 - Care and Maintenance
Cancelling a measurement
A measurement will be cancelled by pressing any buttons during the measurement process. The LCD
display will then show -StoP- and the ABPM 7100 will beep 5 times. This cancellation will be stored in the
measurement value table under Cancel.
Unsuccessful measurement
1. If the display shows errors, reexamine the correct procedure during set-up and positioning of the
device.
2. Dismiss the patient only after a successful manual measurement!
Inform the patient sufficiently in order to explain the situation!
3. Repeat the measurement.
4. If the display still shows errors, repeat the initial operation process.
5. For further troubleshooting measures and faults removal, please refer to the Troubleshooting section.
Note
•
Severe malfunctions are indicated by a continuous beep.
•
In the event of a continuous beep, switch off the device, remove the cuff and inform
your doctor.
P o s: 1 3 / _Steuer ung/= == = == == == SEITEN UMBRU CH== == = == === @ 0\mod_13 999 679 5 7 1 49_ 6.docx @ 48 04 @ @ 1
Care and Maintenance
To ensure the optimal functionality of the ABPM 7100 regular care and maintenance of the unit is required.
Attention
Damage to device
•
Do not open the casing. Once the device is opened, all warranties will lapse.
Cleaning
Cleaning the ABP Monitor and carrying pouch
1. Read the safety instructions carefully and observe them closely before cleaning.
2. Only use a cotton cloth moistened with lukewarm water and mild detergents to clean the ABPM 7100
and the pouch.
Attention
Damage to the ABP Monitor and carrying pouch caused by the use of solvents
•
Do not use strong or solvent-based additives.
•
Ensure that no liquid enters the device.
If liquid does penetrate the device, switch if off immediately and return it to your Welch Allyn
•
specialist for inspection.
Cleaning the cuff sleeve, bladder and tubing
1. Read the safety instructions carefully and observe them closely before cleaning.
2. Before washing, carefully remove the bladder and tubing from the cuff sleeve.
3. When cleaning the cuff sleeve, bladder and tubing, use only mild detergents in lukewarm water without
fabric softener.
Attention
Damage to the cuff sleeve during washing
•
Always close the Velcro strip before washing!
•
It is possible to wash the cuff sleeve in the washing machine at max. 30°C. Do not spin.
Do not use fabric softeners or other washing aids (e.g. hygiene rinses, textile deodorants). These
•
agents can leave behind residues and damage the material.
•
The cuff sleeve is not suitable for drying in a dryer.
Disinfection
Caution
Intolerances caused by the use of disinfectants: Some patients display intolerance (e.g. allergies)
to disinfectants or their components.
•
Never use disinfectants which leave residues on the product or which are not suitable for contact
with the skin.
•
Carefully wash the cuff to remove residues.
Attention
Damage to the cuff sleeve, bladder and tubing caused by disinfectants
Do not submerge the cuff sleeve in disinfectants.
•
Avoid disinfecting the cuff bladder and connected rubber tubing.
•
The bladder and tubing can be damaged by disinfectants. Wipe down the bladder with lukewarm
•
water and add a mild detergent, if necessary.
Ensure that no liquid enters the tube opening.
•
The user (doctor) decides whether and when the ABP Monitor and the cuff sleeve should be disinfected for
hygienic reasons (e.g. after every use).
The following agents are recommended for disinfecting the ABP Monitor and the cuff sleeve:
Terralin Liquid (Manufacturer: Schülke & Mayr)
•
Isopropyl alcohol (70%)
•
For full effectiveness, moisten the ABPM 7100 and cuff sleeve with the disinfectant for at least 5 minutes.
The use of disinfectants not recommended by Welch Allyn shall render the user responsible for proof of safe
application.
Note
It is imperative that you observe the manufacturer's information regarding the use of these
products. Allow the agents to dry off completely.
Maintenance plan
Attention
Damage to device
•
Do not open the casing. Once the device is opened, all warranties will lapse.
Weekly Maintenance
Analysis review:
1. Review the print-out of your measurement analysis for:
•
Correctly entered times and intervals in accordance with the log.
•
Times of day/night transitions.
Correct standard values (nocturnal decrease).
•
2. Check the device, cuff and the cuff tubing for superficial soiling and clean it as specified in the
Cleaning section.
3. Check the cuff and the cuff tubing for superficial damage. In the event of damages return it to your
Welch Allyn specialist for inspection.
Checking battery voltage:
Always use fully charged or new batteries.
The battery voltage appears on the display of the ABPM 7100 for approximately 3 seconds after the device is
switched on. The battery voltage must be at least 2.6 volt to ensure a 24-hour measurement.
Care and Maintenance - 22