25 - Troubleshooting
Error symptom
Possible cause
The patient displays severe
arrhythmia.
Err 1
The arm was moved during
measurement.
Insufficient valid pulse rate detected. Place the cuff on your arm again.
The arm was moved during
measurement.
Err 2
Cuff does not fit the arm snugly.
Blood pressure beyond the
measurement range.
Err 3
Strong arm movements.
Problems with the pneumatics.
Data transmission cable incorrectly
inserted in ABP Monitor.
Pins in the plug of the data
Err 4
transmission cable are mechanically
damaged.
Measurement value was not
correctly transmitted.
Power pack or battery voltage too
low.
Err 5
Power packs or batteries are
defective.
bAtt
Battery contacts are corroded.
Build-up 25 fair.
Err 6 +
Possible continuous
Blood pressure cuff incorrectly
alarm until a button is
connected.
pressed.
Leaky points in the cuff or the cuff
tubing.
The memory of the blood pressure
measurement device is full. (a
Err 7
maximum of 300 measurements and
events can be stored)
Measurement cancelled with a
Err 8
pressed button.
Residual pressure inside the cuff
Err 9 +
Possible continuous
Zero point comparison was
alarm until a button is
unsuccessful.
pressed.
Remedy
ABP Monitor not applicable.
Keep the arm still during measurement.
Keep the arm still during measurement.
Check the seating of the cuff and that of the
device.
Permanent notification renders the ABP
Monitor unsuitable for the patient.
Keep the arm still during measurement.
If the error persists permanently, send the
device to your Welch Allyn specialist.
Insert the cable into the ABP Monitor
correctly.
Check the plug to see whether the pins on
the inside are damaged. If they are, contact
your Welch Allyn specialist.
Restart the transmission.
Replace the power packs or batteries.
The power pack or battery voltage is correct
but "bAtt" is displayed during cuff inflation.
Replace the power packs.
Clean the battery contacts with a cotton
cloth and a little alcohol.
Check the cuff for a build-up of air or a kink
in the tubing. If the cuff tubing is kinked,
straighten the tubing. Otherwise send the
device in for inspection immediately.
Connect the cuff to the device.
If necessary, replace the cuff.
Delete the data in the ABP Monitor but
ensure that the data was stored on your PC
first.
Wait for the cuff to deflate completely.
Send the device to your specialist for
inspection immediately or directly to your
Welch Allyn specialist.
Error symptom
Possible cause
Severe error caused by accumulated
Err 10 +
pressure outside the measurement
Continuous alarm
process.
until a button is
These error messages all show a
pressed.
severe error in the program code.
The analysis unit
does not react to data
Data transmission cable not correctly
transmission but the
inserted in the PC. (also refer to Err 4)
display shows "co".
The ABPM 7100
measures every two
Log 9 is set in the ABPM 7100.
minutes.
The desired log
The last patient's measurement
cannot be set with the
values are still contained in the
button combination.
memory.
The battery packs or batteries were
incorrectly inserted.
The ABP Monitor
Power pack or battery voltage too low. Replace the power packs or batteries.
cannot be switched
on.
Defective display.
An error occurs during
The cuff size is not suitable for the
the first
patient's arm circumference.
measurement.
Communication error ABPM 7100 Bluetooth Interface
Error symptom
Possible cause
®
Code 1
Bluetooth
interface of the ABPM 7100 was not
started correctly.
Possible hardware fault.
®
Bluetooth
interface of the ABPM 7100 could
not be configured correctly.
Code 2
(Communication problem between ABPM 7100
®
and the Bluetooth
module.)
®
The status of the Bluetooth
interface of the
ABPM 7100 could not be determined.
Code 3
(Communication problem between ABPM 7100
and the Bluetooth® module.)
®
The Bluetooth
interface of the ABPM 7100 has
Code 4
not yet been paired with the analysis software.
®
The Bluetooth
interface of the ABPM 7100
Code 5
could not connect to the Bluetooth dongle on
the computer.
The measurement value memory of the ABPM
Code 6
7100 contains unsent blood pressure values.
The ABPM 7100 is paired with a cell phone or
GSM modem, which is technically incapable of
Code 7
transmitting measurement values, is outside the
network range or is incorrectly configured.
Troubleshooting - 26
Remedy
Send the device to your specialist for
inspection and repair immediately or
directly to your Welch Allyn specialist.
Check whether the 9-pin plug of the data
transmission cable is securely seated in
the device's interface socket. (also refer to
Err 4)
Set log 1 or 2.
Delete the data in the ABP Monitor but
ensure that the data was stored first.
Reinsert either power packs or batteries
and ensure correct polarity.
Send the device to your specialist for
repair or directly to your Welch Allyn
specialist.
Measure the patient's arm circumference
and compare this with the imprint on the
cuff. You may require a different cuff size.
Remedy
Send the device to your Welch Allyn
specialist for inspection.
Try again. If the error persists, send
the device to your Welch Allyn
specialist for inspection.
Try again. If the error persists, send
the device to your Welch Allyn
specialist for inspection.
®
Reconnect the device via Bluetooth
.
Try again. If the error persists, send
the device to your Welch Allyn
specialist for inspection.
These will be sent once further
measurements have been performed.
Try again. If the error persists, contact
your Welch Allyn specialist.