9 - Product description
Audible signals
Individual or multiple beeps of audible signals are used. The following table explains the meaning of the
beeps:
1 beep
Switching ON/OFF
Starting and ending a measurement (except at night intervals)
Removal of the interface cable
Establishing and ending Bluetooth® communication
Measurement errors
3 beeps
System errors
Continuous beeps
Severe system errors (e.g. cuff pressure is higher than 15 mmHg for longer than
10 seconds outside the measurement)
Combined beeps
Manual deletion of measurement, 1 beep followed by 5 beeps 2 seconds later
Cuff connection
The cuff connection is located at the top of the ABPM 7100 casing.
The cuff is connected to the ABPM 7100 via a metal connector.
Attention
Measurement errors
The cuff connection must always engage with an audible "CLICK". A poor connection between the
ABPM 7100 and cuff will result in measurement errors.
The Arm Cuff
1
The arm cuff
2
Air tube
3
Air tube connection
PC Interface Cable
In order to read data from the ABPM 7100, the interface cable must be connected to an USB slot on a PC.
PC Interface Cable Port
The connecting port for the PC interface cable is located at the bottom of the ABPM 7100 casing.
The red dot on the plug must align with the red dot on the port before plugging.
To disconnect, pull on the knurled ring of the connector.
Connecting the ABPM 7100 to the PC
To transfer the data from the ABPM 7100, ensure that the interface cable is connected correctly to an USB
port on the PC and to the interface cable port on the device.

Technical Data

Measurement pressure range:
Systolic
Diastolic
Accuracy:
+/- 3 mmHg in display range
Static pressure range:
0 to 300 mmHg
Pulse range:
30 to 240 beats per minute
Procedure:
oscillometric
Measurement intervals:
0, 1, 2, 3, 4, 5, 6, 10, 12, 15, 20 or 30 measurements per hour
Measurement logs:
4 adjustable interval groups
Memory capacity:
300 Measurements
Battery capacity:
> 300 Measurements
Operating temperatures:
+5°C to +40°C
Operating humidity:
15% to 93%
Storage environment:
-25°C to +70°C and 15% to 93% humidity
Dimensions:
121 x 80 x 33 mm
Weight:
approx. 220 g excluding batteries
2 Ni-MH batteries with 1,2 V each and min. 1500 mAh (AA, Mignon) or
Power supply:
2 Alkali 1,5 V batteries (AA, Mignon, LR6)
USB-Interface cable
Interfaces:
Bluetooth® (Class 1 / 100 m and max. 100 mW with 2,402 GHz to
2,480 GHz) only available with optional HMS software
Expected operational device life:
5 Years
Expected operational cuff life:
6 Months
Product description - 10
60 to 290 mmHg
30 to 195 mmHg