Descargar Imprimir esta página

Specifications; Safety Precautions; Installation; Controls - CORNING LSE Manual De Instrucciones

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 18
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.

1.0 Specifications

Temperature range
Temperature display resolution
Temperature uniformity
Temperature accuracy
Temperature controller
Timer
Operating altitude
Control
I/O
Dimensions (W x D x H)
Weight
Electrical requirements
D1100
115V
D1100
230V
D1200
115V
D1200
230V

2.0 Safety Precautions

Do not use this product in an explosive environment .
Do not use in the presence of flammable or combustible material .
Do not heat substances that react violently when heated .
Do not touch block when hot or when unit is heating
(use block lifter) .
Do not touch area around block or block well when unit is hot .
Do not spill liquids into the well area or into the unit side
vent holes .
Connect unit only to a properly grounded outlet .
Caution: HOT Surfaces that can burn or cause injury.
Use of this product in any manner not specified by the manufacturer or
modification of the product may cause injury and/or may void the warranty .
2
Ambient +5°C to 150°C
0 .1°C 4 digit LED
± 0 .2°C (at 37°C in block)
± 0 .3°C
Microprocessor – user calibratable
1 to 999 minutes in 1 minute
increments or continuous
Not to exceed 2,000 meters
PI microprocessor controller
RS232 unidirectional
7 .9 x 10 .4 x 3 .3 in . (20 x 26 .5 x 8 .3 cm)
4 .8 lbs ./2 .2 kg
50/60Hz, 0 .97A, fuse 1 .6 AT
50/60Hz, 1 .48A, fuse 0 .63 AT
50/60Hz, 1 .45A, fuse 2 .0 AT
50/60Hz, 0 .72A, fuse 1 .0 AT

3.0 Installation

Upon unpacking the Corning® LSE™ Digital Dry Bath, inspect for damages .
Shipping damage is the responsibility of the carrier . Inspect that the
following are present: instruction manual, power cord, block lifter .
Select a location that is dry and not subject to drafts or moving air from
heating or air conditioning vents, or air blown by other equipment . Place
the unit on a flat, non-flammable surface . Allow sufficient room around the
unit for access and cooling . Six inches minimum on all sides is suggested .
Plug the unit into a properly grounded outlet . By using the block lifter,
insert the sample block into the well . The unit is now ready for use .

4.0 Controls

On/Off rocker switch, behind the unit: turns primary power on and off
Start/Stop LED, red: starts or stops the unit, illuminates when unit is in
heating mode and off in temperature set mode
Set Temp LED, red: illuminates when unit is in temperature set mode and
off when in heating mode
Heating LED, red: illuminates when unit is in heating mode and unit is
actually applying heat to the block, on continuously during heat-up and
cycles on and off when the unit is at the set temperature
Timer LED, green: sets or selects timer mode, illuminates while running in
timer mode
Up arrow key: raises set temperature or time when unit is in set mode
Down arrow key: lowers set temperature or time when unit is in set mode
Mode key: switches the unit between heating mode, set temperature mode,
or timer mode
3

Publicidad

Solución de problemas

loading