GCE Zen-Olite RS - 00600 Manual Del Usuario página 34

Ocultar thumbs Ver también para Zen-Olite RS - 00600:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 150
15.1.2. Guidance and Manufacturer's Declaration: Electromagnetic Immunity
Zen-O lite™ is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the concentrator should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge
EN
(ESD)
IEC 61000-4-2
Electric fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interrup-
tions, and volt-
age variations
on power
supply
input lines
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
IEC 60601 test
Compliance
level
± 2, 4, 6, 8kV
± 2, 4, 6, 8kV
contact
contact
± 2, 4, 8, 10, 15kV
± 2, 4, 8, 10, 15kV
air
air
± 2kV for power
± 2kV for power
supply lines
supply lines
± 1kV for input/
± 1kV for input/
output lines
output lines
± 1kV differential
± 1kV differential
mode
mode
± 2kV common
± 2kV common
mode
mode
< 5% UT
< 5% UT
(>95% dip in UT)
(>95% dip in UT)
for 0.5 cycle 40%
for 0.5 cycle
UT
40% UT
(60% dip in UT)
(60% dip in UT)
for 5 cycles
for 5 cycles
70% UT
70% UT
(30% dip in UT)
(30% dip in UT)
for 25 cycles
for 25 cycles
<5% UT
<5% UT
(>95% dip in UT)
(>95% dip in UT)
for 5 sec
for 5 sec
30 A/m
30 A/m
Electromagnetic environment/
level
Floors should be wood, concrete, or
ceramic tile.
If floors are covered with synthet-
ic material, the relative humidity
should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should
be that of a typical commercial or
hospital environment. If the user of
the Zen-O  lite™ Portable Oxygen
Concentrator required continued
operation during power main inter-
ruptions, it is recommended that
the concentrator be powered from
an uninterruptible power supply or
battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical commercial or hospital
environment.
32/292
guidance

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido