7.2
Initial startup
WARNINGS
For use only after instruction by medically trained and qualified persons who have been
adequately trained in the use of Thopaz.
CAUTIONS
Before the first application in accordance with the intended use, an inspection
according to IEC 62353 (see Appendix B) is recommended.
7.2a
Checks before use
– Check Thopaz system before use for damage of the power cord or plug, obvious
device damage or safety defects and proper functioning of the device.
– Check for completeness and general condition of Thopaz delivery package.
7.2b
Initial startup
procedure
20 |
Installation
1.
2.
Important: Do not yet connect the patient to the system.
3.
1.1 Select plug.
1.2 Attach plug to the mains
adapter
077.0148.
2.1 Connect Thopaz to mains
by mains adapter
077.0148 or by docking
station
079.0037.
3.1 Charge for approximately
3 hours or until the [
symbol appears in the
display.
3.2 Press [
] to switch
Thopaz on. A selftest
starts.
]