3 Check the display for one of the following visual symbols.
Visual
Symbol
displayed
Blinking!
+
acoustic
signal!
200_4169_ifu_invia_motion_patient_EN_ES_FR.indd 25
What
What do I have to do?
does
this
symbol
mean?
The
Mute the acoustic signal by pressing the
canister
button.
is full
1. Wash hands.
2. Switch the Invia Motion NPWT Pump off
by pressing the
Press only once.
3. Disconnect the Invia Motion Tubing
Connector from the wound tubing
connector by separating the pump from
the patient tubing. Have a paper towel
or tissue avai l able to cover the tip of the
used tubing to protect against wound
fluid leakage.
4. Remove the pump from the carrying
case. Remove the full Canister/Tubing
Set by rotating the canister to the left.
Discard the used Canister/Tubing Set
according to instructions provided by
your healthcare professional.
5. Wash hands.
button for 3 seconds.
➞
25
17/02/15 12:08