ENGLISH
8
Depth: 273mm
Weight: 1790g
Shipping Weight: 2050g
CAUTION
Always check the device before you use it.
Users must use this product under physician's directions.
ALWAYS unplug this product IMMEDIATELY after using.
DO NOT use while bathing.
DO NOT place or store product where it can fall or be pulled into a tub or sink.
DO NOT reach for product that has fallen into water or other liquid. Unplug immediately.
DO NOT place or drop the product into water or other liquid.
This product should NEVER be left unattended when plugged in.
Close supervision is necessary when this product is used by, on, or near CHILDREN AND
PHYSICALLY DISABLED.
Tie up user's long hair if necessary.
DO NOT use attachments not recommended by the manufacturer. Choose attachments from
the brands and specifications recommended by the manufacturer.
NEVER operate this product with a damaged cord or plug, if it is not working properly, if it has
been dropped or damaged, or dropped into water; call Qualified Service Personnel for
examination and repair.
Keep the cord away from HEATED or HOT surfaces.
NEVER use while sleeping or drowsy.
NEVER block the air openings of the product or place it on a soft surface, such as a bed or
couch, where the air openings may be blocked.
Keep the air openings free from lint and hair. Blocked openings may cause the unit to shut
down. Contact dealer immediately.
DO NOT use outdoors or operate where other aerosol (spray) products are being used or
where oxygen is being administered in a confined area.
Do not disassemble in case of the risk of electric shock. Refer servicing to qualified service
personnel of dealer in time.
Do not use nebulizer in anaesthetic breathing systems or lung ventilator breathing systems.
If the unit shuts down frequently, you may have a low voltage situation. Low voltage can also be
suspected when:
- The motor does not get up to full power or speed.
- Home fuses or circuit breakers activate when starting compressor.
- Lights dim or remain dim when compressor is started.
- Other motor-operated appliances fail to operate properly or too many motor-operated appliances
are on same circuit.
The motor of this compressor has a thermal overload protector. If the motor overheated, the
overload protector will shut off the motor. If the unit shuts down and cannot be restarted, the
motor must be replaced. Call your dealer immediately.
Only authorized or factory-trained personnel should perform preventative maintenance.
OPERATION
Your nebulizer package includes a nebulizer, mouthpiece, adaptor, flexi-tubing, air tubing, adult
facemask and child facemask. These items are sold as a package only.
NOTE: Before initial operation, nebulizer should be cleaned. Follow the suggested Cleaning
Instructions in this Owner's Manual.