Gima OXY-4 Manual De Uso página 25

Ocultar thumbs Ver también para OXY-4:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 45
25
B: Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3V/m.
Immunity test
Conducted RF
3 Vrms
IEC 61000-4-6
da 150 kHz 80
MHz
Radiated RF
IEC 61000-4-3
3 V/m
da 80 MHz 2,5
GHz
IEC 60601 test
Compliance
level
N/A
3 V/m
ENGLISH
Electromagnetic environment
Level
Portable and mobile RF
communications equipment should
be used no closer to any part of this
device, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey ,a
should be less than the compliance
level in each frequency range .b
Interference may occur in the
vicinity of equipment marked with
the following symbol:

Publicidad

Tabla de contenido
loading

Tabla de contenido