6.0 TROUBLE
6.1 ERROR
MESSAGES
If an error message is shown, repeat the step that caused the message. If the error
message is still shown then contact your dealer for support.
ERROR
DESCRIPTION
CODE
ADO
A/D Value too high.
ADL
A/D Value too low.
-----
A/D count is not correct
when turning the
balance on.
Unit-E
Unit error.
Low battery indicator. Batteries may be flat.
FAIL
Calibration failure.
Unstable – machine
cannot get a stable
reading.
No Power when turning
on.
© Adam Equipment 2016
SHOOTING
POSSIBLE CAUSES
Weight on the pan when turning
on.
Improper calibration of the
balance.
Damaged load cell.
Damaged Electronics.
Pan is not installed when turning
on.
Calibration not correct.
Damaged load cell.
Load cell damaged.
Electronics damaged.
Incorrect mass used to calibrate.
The user calibration is not within
5% of factory calibration.
Possible damaged to loadcell.
Possible damage to the Loadcell. Make sure the balance is on a
Battery may be flat. DC adapter
may not be working.
14
Solutions
Remove any weight from the
stainless steel pan.
Recalibrate.
Install pan and power back
on.
Recalibrate.
Remove any weight from the
stainless steel pan and try
rebooting. Try calibrating the
balance.
Charge the battery.
Try to recalibrate, check that
the weight used to calibrate
matches that of the value of
the balance.
flat surface and away from
vibration.
Change the batteries. Or use
the DC adapter
308669539revB6-Mar17