Table 3-2. Troubleshooting Guidelines
PROBLEM
POSSIBLE CAUSE
Alarm Test Failure
QuickCheck Station alarm chamber
plugged with dirt or debris
Alarm nonfunctional
Instrument LEDs
LEDs nonfunctional
activated, but
LEDs do not light
during Alarm Test
Vibrating alarm
Vibrator Alarm non-functional
is activated on
instrument, but it
does not vibrate
during Alarm test
Gas Test Failure
No gas in calibration cylinder
Instrument requires calibration
Calibration cylinder tubing improperly installed
QuickCheck station gas delivery outlet
plugged with dirt
Altair Instrument inlet is blocked
Wrong calibration cylinder
Wrong calibration gas concentration
Regulator closed
Expired Cylinder
Incorrect regulator
QuickCheck not
QuickCheck Station not powered
responding to
Altair instrument
Altair instrument button pressed for less than
one second prior to instrument placement in
QuickCheck Station
Instrument not fully seated
No checkmark
Instrument removed prior to test completion
appears on Altair
instrument after
a bump test
SOLUTION
Check for contamination in alarm chamber on Quick-
Check Station and in Altair Detector; clean as needed
Press instrument TEST button; if no sound occurs
during alarm check, contact MSA for replacement
Press instrument TEST button; if LEDs are
nonfunctional, contact MSA for replacement
Press instrument TEST button; if vibrator alarm is
nonfunctional, contact MSA for replacement
Check pressure gage on calibration cylinder
Calibrate instrument per MSA recommendations
Verify tubing is fully connected to QuickCheck
Station fitting
Clean gas delivery outlet
Check for blockage and remove if necessary
Verify that calibration cylinder is the same gas type as
the Altair instrument
Check manual and calibration cylinder for
proper concentration
Ensure regulator is opened or turned ON before
placing the instrument into the QuickCheck Station
Check expiration date of calibration cylinder;
replace if necessary
Verify that QuickCheck Station is used with
the recommended regulator
Check power connections to QuickCheck Station
Remove instrument; press and hold TEST button for
one full second before placing instrument into
QuickCheck Station
Fully seat instrument in QuickCheck Station
Re-test instrument
3-3