- Rapid transients in accordance with IEC 1000-4-4 (1995):
. test voltage: 1 kV level 2, aptitude criteria B
- Magnetic fields at the frequency of the network to IEC 1000-4-8 (1995):
. with a max distortion of 0.5 A: 30 A/m 50 Hz level 4, aptitude criteria A
Emissions in accordance with EN 50081-1:
- Conducted and radiated emission through the case to EN 55022 (1994):
class B
For maintenance, use only specified spare parts. The manu-
facturer will not be held responsible for any accident occurring
following a repair done other than by its After Sales Service
or approved repairers.

5.1/ REPLACING THE BATTERY

- Completely disconnect the clamp from the circuit to be measured and
the measurement instrument it is connected to.
- Unscrew the tool release screw holding the cover of the battery com-
partment.
- Replace the 9 V battery (type 6LF22, 6LR61 or NEDA 1604).
- Replace the cover of the battery compartment.

5.2/ CLEANING

Keep the jaw faces and mechanism perfectly clean.
The body of the clamp should be cleaned with a cloth moistened with
soapy water.
Rinse with a cloth moistened with clean water.
Never expose the clamp to running water.

5.3/ CALIBRATION

It is essential that all measuring instruments are regularly
calibrated.
For checking and calibration of your instrument, please contact our
accredited laboratories (list on request) or the Chauvin Arnoux subsidiary
or Agent in your country.
5.4/ REPAIRS
Repairs under or out of guarantee: please return the product to your
distributor.