Mettler Toledo IND780 Guia De Instalacion página 65

Ocultar thumbs Ver también para IND780:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 97
4. Twist the home run cable's inner drain wire together with the black and green ground wires and
2.7.4.3.4.
Open Collector Output
The POWERCELL PDX option board provides a single TTL-compatible open collector output (pin 1
and pin 2), with current sinking components that can handle from 5 to 30 VDC signals at a
maximum of 35 mA current. A solid state relay or OPTO 22 is typically connected to buffer the
output to a 120 or 220 VAC signal. The voltage source must be a Class 2 marked or a Limited
Power Source (LPS)-compliant power supply.
This output is reserved for use with the scale's material transfer target functionality. It provides the
feed output signal associated with the active target for a POWERCELL PDX scale assigned as scale
1.
64057253
|
14
|
08/2014
attach them to the external ground stud, as shown in Figure 2-67.
Inner drain wire
and ground wires
Panel enclosure
grounding stud
Figure 2-67: Completed POWERCELL PDX Home Run Power Ground Connection,
Panel Mount Enclosure
METTLER TOLEDO IND780 Installation Guide
Ground wire
connected to pin
7 of PDX option
board
2-41

Publicidad

Tabla de contenido
loading

Tabla de contenido