Power-On And Remote Operation Default Settings - Metek Sorensen XEL serie Manual De Operación

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Power-on and Remote Operation Default Settings

The following instrument status values are set at power on:
Status Byte Register
Service Request Enable Register †
Standard Event Status Register
Standard Event Status Enable Register †
Execution Error Register
Query Error Register
Parallel Poll Enable Register †
† Registers marked thus are specific to the GPIB section of the instrument and are of limited use via
other interfaces.
The instrument will be in local state with the front panel controls active (providing rear panel
switches CV and CC are set to Local.
The instrument parameters at power-on are the same as at last switch off with the exception of the
output status. By default this is always off at power on but the user may change this from the front
panel to the same at power on as at switch off, see Switching On and Power-on Conditions
paragraph in the Manual Operation section.
The *RST (reset) interface command resets the instrument to the Remote Operation Default
settings.
Remote Operation Default settings are:
V
=1V, I
=10mA, DeltaV=100mV, DeltaI=1mA, Lock cancelled, V-Span cancelled, Low Range
out
out
cancelled, Meter Average cancelled, OVP & OCP at 5% above instrument range maximums.
Remote interface settings, stored set-ups, Vmin/Vmax values and Output state at power-on setting
are unchanged by *RST.
= 0
= 0
= 128 (pon bit set)
= 0
= 0
= 0
= 0
31

Publicidad

Tabla de contenido
loading

Productos relacionados para Metek Sorensen XEL serie

Este manual también es adecuado para:

Sorensen xel 120Sorensen xel 250

Tabla de contenido