Adjust Gain (Sensitivity Selection): Sensitivity is to be selected by pressing key
User can choose 2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV or 40mm/mV.
Adjust the speed: Use the key of speed adjusting
6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s. Auto-record and Rhythm record can not
support 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s when printing.
Shift the filter: Use the key of filter selection
non-filter,AC,EMG,DFT,AC+EMG,AC+DFT,EMG+DFT,AC+EMG+DFT.
Display the calibration: Use the key of "1 mV"
screen.
Print/ finish the print: Use the key of "print"
printing operation.
Automatic mode: After starting printing, the system will print and store synchronic twelve leads
waveform automatically. The length decided by the setting item in the printing option. And also
according to the setting item, print out the data and result which analysed automatically and the
system can finish the printing automatically.
Manual mode: After starting printing, the user should print out the real waveform by shifting the
group of leads. That means the ECG waveform printing under the manual mode is
non-synchronic, and cannot be saved. And the user should end the printing by press the key
again.
During printing, the printing state includes:
Display content
Process...
Waiting...
No Paper.
Print Timeout.
ECG Timeout
Low Power
Note: Please print after the ECG was displayed in the screen .
On this interface, press the button
setting interface:
User Manual
In the process of printing
In the process of ending printing
Lack of paper, the user should restart the operation after filling
papers.
The connection between system and printing sub-system broke.
The connection between system and sampling sub-system broke.
Low power, the system can not start the print work.
on key board, the system will enter the shortcut
20
to change the speed :5mm/s,
to shift between
to display the marker of 1 mV on the
on the keyboard, then can start or stop a
Remark
.