Data Logging
Session Log
•
The ALTAIR Pro Single Gas instrument has the capability of
recording 50 of the most recent events.
To transfer data to a pc, hold the TEST button for one second.
•
The instrument pages begin to display.
•
Put the instrument in IR mode (see FIGURE 2-5).
•
Events will be transferred to the PC during this sequence, if:
•
the top of the instrument is pointed toward the optional IR
Receiving device
•
a PC is running MSA FiveStar Link Software (P/N 710946)
•
"CONNECT" is pressed in the PC FiveStar LINK Software
package.
•
The following events are recorded:
•
Alarm - Alarm Type - Alarm Value - Time/Date
•
Alarm Clear - Alarm Type - Alarm Value - Time/Date
•
Cal (Pass/Fail) - Time/Date
•
Bump (Pass/Fail) - Time/Date
•
Error - Error Type (See Error List) - Time/Date
Periodic Log
•
Peak gas readings are periodically logged, based on user settings
via the PC. (Default is one sample every three minutes.)
•
See TABLE 6-3 for typical performance.
NOTES:
•
The time and date is based on the PC time and date.
Ensure the PC is correct before communicating with the
instrument.
•
Changing the instrument battery can result in lost time in
the Data log.
•
After changing batteries, verify the Time and Date with
your PC.
2-17