Anexo A: Informações sobre compatibilidade eletromagnética
GUIDANCE AND MANUFACTURER'S DECLARATION- ELECTROMAGNETIC IMMUNITY:
This device is intended for use in the electromagnetic environment specified below. The user of this device should make
sure it is used in such an environment.
Basic EMC
Immunity Test Levels
standard
HOME HEALTHCARE
ENVIRONMENT
Electrostatic
±8kV contact
Discharge (ESD)
±15kV air
IEC61000-4-2
Electrical fast
±2kV for power supply line
transient/ burst
±1kV for input/output line
IEC61000-4-4
Surge
± 1 kV line(s) to
IEC61000-4-5
line(s)
± 2 kV line(s) to earth
Voltage dips, short
Voltage Dips:
interruptions and
i) 100% reduction for 0.5
voltage variations
period,
on power supply
ii) 100% reduction for 1
input lines
period,
IEC61000-4-11
iii) 30% reduction for 25/30
period,
Voltage Interruptions:
100% reduction for 250/300
period
Power frequency
30 A/m
(50/60Hz)
magnetic field
IEC61000-4-8
Conducted RF
3 Vrms
IEC 61000-4-6
0,15 MHz – 80 MHz
6 Vrms in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
Regressar ao Índice
18
Compliance Levels
±8kV contact
±15kV air
±2kV for power supply
line
±1kV for input/output
line
± 1 kV line(s) to
line(s)
230V (U
)
(1)
T
Voltage Dips:
i) 100% reduction for 0.5
period,
ii) 100% reduction for 1
period,
iii) 30% reduction for
25/30 period,
Voltage Interruptions:
100% reduction for
250/300 period
30 A/m
6Vrms
(4)
476005-5210 V1.6
Electromagnetic
Environment-Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of this
device requires continued operation
during power mains interruptions, it
is recommended that the device be
powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
O equipamento de comunicação portátil
e RF móvel não deve ser usado próximo
de qualquer parte deste dispositivo,
incluindo os cabos, a uma distância
inferior à recomendada e calculada
segundo a equação aplicável à
frequência do transmissor.