Descargar Imprimir esta página

Emc Information (Electromagnetic Compatibility Information) - NSK VIVA MATE G5 Manual De Instrucciones

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 69
16

EMC Information (Electromagnetic Compatibility Information)

Guidance and manufacturer's declaration - Electromagnetic Emissions
VIVA MATE G5 is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such
an environment.
Emissions test
RF emissions
CISPR11/ EN55011
RF emissions
CISPR11/EN55011
Harmonic emissions
EN/IEC61000-3-2
Voltage fluctuations/flicker emissions
EN/IEC 61000-3-3
Guidance and manufacturer's declaration - Electromagnetic Immunity
VIVA MATE G5 is intended for use in the electromagnetic environment specified below. The customer or the user of VIVA MATE G5 should assure that it is used in
such an environment.
Immunity test
IEC60601 test level
Electrostatic discharge (ESD)
±(2,4,6)kV contact
EN/IEC61000-4-2
±(2,4,8)kV air
Electrical fast transient/burst
±2kV for power supply lines
EN/IEC61000-4-4
±1kV for input/output lines
Surge
±1kV line(s) to line(s)
EN/IEC61000-4-5
±2kV line(s) to earth
Voltage dips, short
<5% Ut (>95% dip in Ut)
interruptions and voltage
for 0.5 cycles
variations on power supply
40% Ut (60% dip in Ut)
input lines
for 5 cycles
EN/IEC61000-4-11
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec
Power frequency
3 A/m
(50/60Hz)magnetic field
EN/IEC61000-4-8
NOTE: 'Ut' is the AC mains voltage prior to application of the test level.
Compliance
Electromagnetic environment - guidance
Group 1
VIVA MATE G5 uses RF energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.
VIVA MATE G5 is suitable for use in all establishments, including domestic establishments and
Class B
those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Not Applicable
Not Applicable
Compliance level
±(2,4,6)kV contact
±(2,4,8)kV air
±2kV for power supply lines
±1kV for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% Ut (>95% dip in Ut)
for 0.5 cycles
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec
3 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment. If the user of VIVA MATE G5 requires
continued operation during power mains interruptions, it is
recommended that VIVA MATE G5 be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.

Publicidad

loading