OPERATION

1.
Connect the lamp to the power supply.
2.
Connect the power supply to line power.
3.
Press the power button on the rear of the lamp housing to turn on the UV LEDs.
4.
The lamp will immediately be at full intensity and will stabilize within 5 minutes.
5.
To turn off, press the power button on the rear of the lamp housing.
NOTE: The EV6000 is designed for continuous use; operating the lamp only when required
will increase the lifespan of the UV LEDs.

MAINTENANCE

Check UV-A intensity at regular intervals using a calibrated UV-A Meter (320-400 nm range)
UV-A intensity readings should be taken with the lamp positioned 15 in / 38 cm from the
lamp filter to the face of the sensor
Use mild soap and a damp cloth to clean the lamp housing as needed
Do not spray lamp or power supply or immerse any part in liquid

SERVICE

CAUTION
To assure product safety and reliability, repairs should be performed by
Magnaflux Authorized Service Center using original parts.
WARNING
The LED assembly is static sensitive and can be permanently damaged with
improper handling. All replacements of the LED assembly must be performed by a
Magnaflux Authorized Service Center. Unauthorized repairs will void the warranty.
WARNING!
To maintain cTUVus certification, service of model 628785 must be performed by
the factory. Contact Magnaflux Customer Service for more information.
UV FILTER REPLACEMENT
1.
Disconnect the lamp from the power supply.
2.
Remove the protective shade from the front of the lamp.
3.
Using a 3-mm hex key, remove the four screws from the front bezel.
4.
Remove the front bezel, then remove the UV Filter and gasket.
5.
Inspect the gasket for tears or damage. If damage is found, replace the gasket.
6.
Install the new filter, placing the gasket below the filter in the plastic inset tray.
7.
Replace the front bezel and secure with the four hex screws. NOTE: Tighten hex screws
until they stop (little to no force required), then tighten an additional 1/8 turn.
8.
Replace the protective shade.
DISPOSAL
The EV6000 contains no hazardous materials and complies with EU RoHS regulations.
Please contact your local waste disposal authority for instructions on disposal of electrical waste.