Pediatric Reduced Energy Defibrillation Electrodes Instructions for Use
Patient's
Impedance
(ohms)
25
50
75
100
125
Patient's
Impedance
(ohms)
25
50
75
100
125
Patient's
Impedance
(ohms)
25
50
75
100
125
6
Ultra Low Energy (150 VE) – All values are typical
Phase 1
Voltage
Duration
(volts)
(ms)
370
5.8
550
6.5
640
7.0
705
7.4
770
7.8
Low Energy (200 VE) – All values are typical
Phase 1
Voltage
Duration
(volts)
(ms)
430
5.8
630
6.5
745
7.0
790
7.4
855
7.8
High Energy (300 VE) – All values are typical
Phase 1
Voltage
Duration
(volts)
(ms)
500
5.8
700
6.5
820
7.0
920
7.4
960
7.8
Phase 2
Voltage
Duration
(volts)
(ms)
270
3.2
390
3.2
470
3.2
510
3.2
545
3.2
Phase 2
Voltage
Duration
(volts)
(ms)
295
3.2
425
3.2
510
3.2
560
3.2
610
3.2
Phase 2
Voltage
Duration
(volts)
(ms)
380
3.2
520
3.2
620
3.2
680
3.2
720
3.2
Energy
(Joules)
31
35
34
32
29
Energy
(Joules)
41
47
46
43
39
Energy
(Joules)
56
63
62
58
53